| Toxoplasma gondii IgG ELISA |
| |
| BBI T. gondii Performance Panel, IgG results
Toxoplasma gondii performance panel (Cat#PTT201) of FDA regulated Boston Biomedica Inc. (BBI), Milford, MA, USA, was also evaluated using Orgenium T. gondii IgG ELISA test.
|
 |
 |
| Click for table |
| Less |
| |
BBI Sample ID | Abbott
IMx Toxo-IgG Lot:
96433Q100 | BioMerieux VIDAS Toxo-IgG Lot:
960221-0 | Orgenium          T.
gondii IgG     Lot: TXGG070813           Ratio (S/Co)
| Orgenium      T.
gondii IgG Interpretation | | 1 | Positive | Positive | 2.26 |         
Positive | | 2 |       Negative |        
Positive | 0.88 |         
Negative | | 3 | Negative |
        Negative | 0.26 |         
Negative | | 4 | Positive | Positive | 4.18 |         
Positive | | 5 | Positive | Positive | 1.9 | Positive | | 6 | Positive | Positive | 3.62 | Positive | | 7 | Positive | Positive | 1.73 | Positive | | 8 | Positive | Positive | 3.78 | Positive | | 9 | Positive | Positive | 4.77 | Positive | | 10 | Positive | Positive | 1.94 | Positive | | 11 | Positive | Positive | 4.27 | Positive | | 12 | Positive | Positive | 3.30 | Positive | | 13 | Negative | Negative | 0.21 |          
Negative | | 14 | Positive | Positive | 1.67 | Positive | | 15 | Positive | Positive | 2.16 | Positive | | 16 | Positive | Positive | 2.10 | Positive | | 17 | Positive | Positive | 5.19 | Positive | | 18 | Positive | Positive | 1.79 | Positive | | 19 | Positive | Positive | 3.35 | Positive | | 20 | Positive | Positive | 2.09 | Positive | | 21 | Negative | Negative | 0.23 |          
Negative | | 22 | Positive | Positive | 2.80 | Positive | | 23 | Positive | Positive | 5.20 | Positive | | 24 | Positive | Positive | 2.00 | Positive | | 25 | Positive | Positive | 4.24 | Positive |
|
|
| |
| |
| Toxoplasma gondii IgM ELISA |
| |
| BBI T. gondii IgM Performance Panel, IgM results
Toxoplasma gondii performance panel (PTT201) of FDA regulated Boston Biomedica Inc. (BBI), Milford, MA, USA, was also evaluated using Orgenium T. gondii IgM EIA test.
|
 |
 |
| Click for table |
| Less |
| |
Sample ID |
Abbott IMx Lot:  95441Q100 | BioMerieux VIDAS Lot: 951215-0 | Orgenium T. gondii IgM Lot: TXGM070813 Ratio (S/Co) | Orgenium  T. gondii IgM
Interpretation | | 1 | Negative | Negative | 0.42 | Negative | | 2 | Negative |
Negative | 0.26 | Negative | | 3 | Negative |
Negative | 0.30 | Negative | | 4 | Negative |
Negative | 0.31 | Negative | | 5 | Negative |
Negative | 0.84 | Negative | | 6 | Negative |
Negative | 0.33 | Negative | | 7 | Negative |
Negative | 0.15 | Negative | | 8 | Positive |
Positive | 2.14 | Positive | | 9 | Negative |
Negative | 0.34 | Negative | | 10 | Negative |
Negative | 0.29 | Negative | | 11 | Positive |
Positive | 2.08 | Positive | | 12 | Negative |
Negative | 0.37 | Negative | | 13 | Positive |
Positive | 2.18 | Positive | | 14 | Negative |
Negative | 0.56 | Negative | | 15 | Negative |
Negative | 0.24 | Negative | | 16 | Negative |
Negative | 0.60 | Negative | | 17 | Positive |
Positive | 0.9 | *Uncertain | | 18 | Negative |
Negative | 0.20 | Negative | | 19 | Negative |
Negative | 0.46 | Negative | | 20 | Negative |
Negative | 0.20 | Negative | | 21 | Negative |
Negative | 0.40 | Negative | | 22 | Negative |
Negative | 0.31 | Negative | | 23 | Positive |
Positive | 1.56 | Positive | | 24 | Negative |
Negative | 0.25 | Negative | | 25 | Positive |
Positive | 1.50 | Positive |
  *This sample (#17) was found negative with another commercially
available CE marked T. gondii IgM capture assay.
Activity loss of the sample due to freeze-thaw cycles is suspected.
|
|
| |
| |
| Rubella IgG ELISA |
| |
| BBI Rubella IgG Performance Panel
Rubella performance panel (PTR201) of FDA regulated Boston Biomedica Inc. (BBI), Milford, MA, USA, was also evaluated using Orgenium Rubella IgG ELISA test kit.
|
 |
 |
| Click for table |
| Less |
| |
| SAMPLE ID | Abbott-IMX EIA Rubella IgG Lot:
02893M200 | BioWhittaker EIA
Rubella-IgG Lot: 4E2319 | Orgenium Rubella
IgG Lot:
RUBG070813 Ratio (S/Co) | Orgenium Rubella
IgG Interpretation | | 1 | Positive | Positive | 4.72 | Positive | | 2 | Positive | Positive | 5.496 | Positive | | 3 | Positive | Positive | 7.632 | Positive | | 4 | Positive | Positive | 7.808 | Positive | | 5 | Negative | Negative | 0.296 | Negative | | 6 | Positive | Positive | 1.604 | Positive | | 7 | Positive | Positive | 2.556 | Positive | | 8 | Positive | Positive | 2.06 | Positive | | 9 | Positive | Positive | 6.62 | Positive | | 10 | Positive | Positive | 4.172 | Positive | | 11 | Positive | Positive | 2.728 | Positive | | 12 | Positive | Positive | 7.732 | Positive | | 13 | Positive | Positive | 6.484 | Positive | | 14 | Positive | Positive | 1.324 | Positive | | 15 | Positive | Positive | 7.768 | Positive | | 16 | Negative | Negative | 0.448 | Negative | |         17 |
Positive | Positive | 3.916 | Positive | | 18 | Positive | Positive | 2.172 | Positive | | 19 | Positive | Negative | 2.004 | Positive | | 20 | Positive | Positive | 5.244 | Positive | | 21 | Positive | Positive | 3.54 | Positive | | 22 | Positive | Positive | 1.076 | Positive | | 23 | Positive | Positive | 3.392 | Positive | | 24 | Positive | Positive | 1.376 | Positive | | |